Related Experiment Video
Updated: Mar 18, 2026

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Reference Levels for Diagnostic and Interventional X-ray Exposures in Germany: Development, Handling, and Perspective
Hugo de Las Heras Gala1, Augusto Giussani1, Helmut Schlattl1
1External and Internal Dosimetry, Biokinetics, Federal Office for Radiation Protection, Neuherberg, Germany.
Diagnostic reference levels (DRLs) effectively reduce patient radiation exposure from X-ray procedures in Germany. Over nearly 20 years, DRLs have been significantly lowered through technological advancements and optimized practices, enhancing patient protection.
Area of Science:
- Medical Physics
- Radiology
- Radiation Protection
Background:
- Diagnostic and interventional X-ray applications constitute over 95% of man-made radiation exposure in Germany.
- National Diagnostic Reference Levels (DRLs) are crucial for optimizing these procedures and ensuring patient safety.
- Regular adjustments to DRLs are necessary to incorporate advancements in device technology and clinical practices.
Purpose of the Study:
- To present the fundamental concept of Diagnostic Reference Levels (DRLs).
- To detail the methodological development and evolution of DRLs in Germany.
- To highlight the improvements in patient protection resulting from the implementation of DRLs.
Main Methods:
- Summarizing relevant regulations within Germany's new radiation protection legislation.
- Describing the historical development of DRLs in Germany from 2003 to 2022.
- Providing guidance for the practical application of DRLs and facility-specific measures.
Main Results:
- DRLs in Germany have been substantially reduced over nearly two decades.
- Continuous protocol optimization and advanced hardware/software have led to significant dose reductions.
- The scope of X-ray examinations covered by DRLs has expanded, with some obsolete DRLs removed.
Conclusions:
- The consistent implementation of the DRL concept is effective in practice, leading to substantial reductions in patient exposure.
- The positive impact on patient protection justifies the effort and potential investment in updating technology.
- Continuous development of the DRL concept is essential for ongoing optimization of radiation protection.
More Related Videos
Related Concept Videos
X-ray Imaging
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...