Purpose:
To develop a deep vacuum vaporized hydrogen peroxide (VH₂O₂) sterilization method capable of preserving the structural and functional integrity of biocompatible polymeric electrospun mats usable as advanced biocompatible medical implants and devices, with particular focus on the influence of vacuum level during treatment.
Methodology:
Electrospun mats made of poly-L-lactide-co-glycolide (PLGA), poly-L-lactide-co-ε-caprolactone (PLC), and thermoplastic polyurethane (TPU) were prepared and sterilized with VH₂O₂ under low-vacuum (LV, 10 mbar) and high-vacuum (HV, 2 mbar) conditions at 20-50 °C. Scanning electron microscopy (SEM) assessed fiber morphology and pore area, contact angle measurements evaluated surface wettability, Fourier-transform infrared spectroscopy (FTIR) detected chemical modifications, and gel permeation chromatography (GPC) analyzed weight-average molecular weight (Mw) and polydispersity index (PI).
Results:
HV sterilization caused fiber compression across all polymers, increasing fiber diameter and reducing pore area. LV sterilization conditions prevented morphological changes in TPU and PLC, while PLGA remained sensitive under both vacuum levels. PLGA also exhibited increased hydrophilicity due to surface reorganization and transition from Cassie-Baxter to Wenzel-type wetting. FTIR showed no chemical changes, and GPC confirmed stable Mw and PI for all tested materials (PLGA: 25-27 kDa; PLC: 16-18 kDa; TPU: 23-25 kDa).
Conclusion:
Deep vacuum VH₂O₂ sterilization at room temperature is feasible, with vacuum level critical for maintaining scaffold integrity. LV sterilization effectively preserves morphology in PLC and TPU, while HV is also reliable for less sensitive polymers. Chemical and molecular properties remain unaffected, supporting VH₂O₂ sterilization as a gentle and effective promising method for implantable textile-based biomaterials.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration