Representation and performance of CE-marked and in-house in vitro diagnostic tests for routine laboratory parameters
Christoph Buchta1, Andreas Bietenbeck2, Andrea Griesmacher1,3
1Austrian Association for Quality Assurance and Standardization of Medical and Diagnostic Tests (ÖQUASTA), Vienna, Austria.
Objectives:
European legislation requires detailed performance evaluation of in vitro diagnostic medical devices (IVD-MD) by both commercial manufacturers (through the Conformité Européenne (CE) mark) and in-house (IH) IVDs. Little information is currently available on how CE- and IH-IVDs are mapped in laboratory routine operations with respect to spectrum, frequency and performance.
Methods:
External quality assessment (EQA) providers routinely collect information from participants in connection with test results on the IVD-MDs and combinations used. Using data from a national EQA provider, we compared frequencies and performances of commercial and non-commercial laboratory tests reported by laboratories participating in the programmes of an EQA provider. Due to differing, but equally tenable, interpretations regarding the distinction between CE- and IH-IVDs, two possibilities were considered in data analysis.
Results:
Depending on the applied definition of IH-IVDs, a total of 59 or 13/1,763 laboratories (3.5 %/0.74 %) used 170 or 21/5,488 (3.1 %/0.4 %) IH-IVD systems to produce a total of 263 or 21/32845 (0.8 %/0.1 %) results for 63 or 18 of ∼400 analytes (∼15.8 %/∼4.5 %). The EQA performance of IH-IVDs was essentially the same as that of CE-IVDs.
Conclusions:
IH-IVDs accounted for only a minimal proportion of tests. EQA results obtained with them showed no inferiority to those obtained with CE-IVDs. Furthermore, our findings show that EQA data are a useful tool for examining the analytical landscape, possibly even beyond the national level. Studies on the use and performance of IH-IVDs should be conducted based on EQA programmes for laboratory diagnostic areas where their proportion is likely to be higher.
Related Concept Videos
Quality Assurance
Automated Microbial Diagnostics
Serum Laboratory Studies, Stool Test, Breath Test
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
In Vitro Drug Release Testing: Overview, Development and Validation
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...