Value of the Acuros reconstruction CBCT algorithm for conventional (non-HyperSight) Ethos systems
Erik B van Dieren1, Anieck E Ter Braak2, Esther Bosman2
1Medisch Spectrum Twente, Department of Radiotherapy, Enschede, the Netherlands.
Zeitschrift Fur Medizinische Physik
|March 18, 2026
Summary
The Acuros algorithm improves image quality for non-HyperSight Ethos systems (NHES) cone-beam CT (CBCT) in adaptive radiotherapy. This advanced reconstruction enhances contrast and uniformity, leading to better image interpretation for pelvic treatments.
Area of Science:
- Medical Physics
- Radiotherapy Technology
- Image Reconstruction
Background:
- Online adaptive radiotherapy utilizes cone-beam CT (CBCT) to adjust treatment based on daily anatomical changes.
- Varian's Ethos system enables CBCT-based adaptive radiotherapy, but pelvic CBCTs suffer from low contrast.
- Non-HyperSight Ethos Systems (NHES) require optimized image reconstruction for adaptive treatments.
Purpose of the Study:
- To evaluate the added value of the Acuros algorithm for reconstructing pelvic CBCT images on NHES.
- To compare Acuros reconstruction against standard iCBCT reconstruction for phantom and clinical data.
- To assess image quality metrics and observer preference for Acuros-reconstructed CBCTs.
Main Methods:
- Anthropomorphic pelvis phantom CBCTs were acquired bi-weekly, with CT contours propagated.
- Image quality metrics including Signal-to-Noise Ratio (SNR), Contrast-to-Noise Ratio (CNR), uniformity, and resolution were calculated.
- Clinical CBCTs from five prostate cancer patients undergoing adaptive radiotherapy were analyzed, with Acuros reconstruction applied to the pre-treatment CBCTs. Observer scoring was also performed.
Main Results:
- Acuros reconstruction demonstrated superior contrast-to-noise, signal-to-noise, and uniformity compared to iCBCT in phantom studies.
- For clinical CBCTs, Acuros yielded improvements in contrast-to-noise, uniformity, and resolution.
- Observers rated Acuros images as better or similar to iCBCT in 85-95% of cases for the adaptive task.
Conclusions:
- The Acuros algorithm generally outperforms standard reconstruction for pelvic CBCTs on NHES, particularly in phantom studies.
- Clinical evaluation and observer preference indicate that Acuros provides superior or comparable image quality for adaptive radiotherapy.
- The use of Acuros reconstruction is recommended for NHES to enhance adaptive pelvic radiotherapy accuracy.