Single-use optical tracking navigation for external ventricular drain placement: Preclinical 3D-Printed and cadaveric
Daniel De Wilde1, Attill Saemann1, Roman Ramseyer2
1Department of Neurosurgery, University Hospital of Basel, Basel, Switzerland.
Brain & Spine
|March 19, 2026
Summary
A novel single-use optical tracking navigation (OTN) device demonstrated high accuracy for external ventricular drain (EVD) placement in simulated models. This technology offers a user-friendly solution to improve EVD insertion accuracy in neurosurgery.
Area of Science:
- Neurosurgery
- Medical Devices
- Image-Guided Therapy
Background:
- External ventricular drain (EVD) placement is a common neurosurgical procedure.
- Malposition rates for EVDs can be as high as 38%, necessitating improved placement accuracy.
- Current technical adjuncts exist, but simpler, user-friendly navigation solutions are needed.
Purpose of the Study:
- To evaluate the feasibility and accuracy of a novel single-use optical tracking navigation (OTN) system for EVD placement.
- To assess the usability of the OTN device in both 3D-printed and cadaveric head models.
Main Methods:
- The study involved neurosurgeons performing OTN-assisted EVD placements in 9 3D-printed head models and 2 cadaveric heads.
- Accuracy was primarily assessed using the Kakarla grading system.
- Additional outcomes included fluoroscopy-derived tip-to-target distance, procedure duration, and user satisfaction (Likelihood to Recommend - LTR).
Main Results:
- All 11 EVD placements achieved Kakarla grade 1, indicating successful and accurate placement.
- Mean fluoroscopy-derived tip-to-target distances were 13.3 mm (3D-printed) and 22.2 mm (cadaveric).
- The average procedure time was 12.1 minutes, with positive user satisfaction (mean LTR of 7/10).
Conclusions:
- The single-use OTN device facilitates accurate, real-time image-guided EVD placement in simulated environments.
- The device's accuracy appears independent of user experience.
- Further research, including larger cadaveric studies and clinical trials, is warranted to confirm accuracy and clinical utility.