Related Experiment Video
Updated: Mar 21, 2026

Author Spotlight: Using Hyperpolarized Xenon-129 MRI to Study Lung Diseases
Published on: January 5, 2024
Well-designed k-space coverage is important for good MRI denoising
Jiayang Wang1, Justin P Haldar2
1Signal and Image Processing Institute, Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, 3740 McClintock Ave, EEB 400, Los Angeles, CA, 90089, USA. jiayangw@usc.edu.
Objective:
Modern computational MRI denoising approaches are often designed assuming fixed k-space coverage. This contrasts with earlier acquisition-design literature that leveraged k-space coverage modifications (e.g., reducing spatial resolution) to improve SNR. This work investigates whether the performance of modern computational denoising methods can be further enhanced by k-space coverage modifications.
Materials And Methods:
Using realistic simulations of noisy data, k-space coverage and averaging patterns were optimized for two advanced image denoising/reconstruction approaches: parallel imaging with total variation regularization and a U-Net neural network. For reference, comparisons against classical linear filtering/apodization methods were also performed. Performance was quantified using normalized root-mean-squared error (NRMSE) and structural similarity (SSIM) metrics.
Results:
Advanced computational denoising methods can be substantially enhanced, both quantitatively and qualitatively, by reducing the spatial resolution of the acquisition to improve SNR. Indeed, even simple linear filtering/apodization with optimized k-space coverage can rival advanced methods using naive higher-resolution coverage.
Discussion:
Classical acquisition design principles that allow spatial resolution to be traded for SNR enhancement are still very relevant for modern computational denoising techniques. However, the optimization of k-space coverage and denoising/reconstruction methods can also be somewhat confounded because the NRMSE and SSIM metrics have low sensitivity to spatial resolution.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
NMR Spectrometers: Resolution and Error Correction