Intraoperative Degree Misalignment in Toric Intraocular Lens Implantation When Using Conventional Compared to Digital
Rom Kandavel1, Justin Dredge1, Hayden Jackson1
1Colvard-Kandavel Eye Center, Encino, CA, USA.
Purpose:
To evaluate the difference in degrees between a digital marker and a manual marker as measured intraoperatively.
Methods:
This was a prospective, single-site, comparative, observational study. Preoperative biometry and planning was assessed with the Argos SS-OCT biometer. All subjects received both digital and conventional marking. The ToriCAM application was used to mark the patient manually at the bedside. A Mendez marker was used to mark the final axis of the IOL, as calculated by Argos. Digital marking was performed using the VERION image guided system. The primary outcome measure was the difference in degrees between the digital marker and the manual marker as measured intraoperatively. Other outcome measures included IOL rotation, residual astigmatism, absolute prediction error (APE), and monocular visual acuity at 2 months postoperatively.
Results:
A total of 41 eyes of 41 subjects completed the study. At 2 months postoperatively, the absolute difference in degrees between intraoperative digital and manual marking was 6.6 ± 5.2 (range 0.0 to 22.5). Mean absolute IOL rotation in degrees was 3.5 ± 2.9 (range 0.3 to 14.8) at 2 months. Mean postoperative residual astigmatism was 0.24 ± 0.24 D, and mean APE (spherical equivalent) was 0.39 ± 0.27 D. Mean monocular corrected distance visual acuity was 0.02 ± 0.05 logMAR.
Conclusion:
Results suggest good postoperative refractive accuracy was achieved using Argos combined with a digital marker microscope. The manual marking was significantly misaligned when compared to the intended digital marking axis, which may cause significant postoperative residual astigmatism.
More Related Videos
05:54Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Errors in Taping
Common Leveling Mistakes and Errors