Related Experiment Video
Updated: Jul 10, 2026

3D Kinematic Analysis for the Functional Evaluation in the Rat Model of Sciatic Nerve Crush Injury
Published on: February 12, 2020
Sacroiliac fixation in pediatric patients: stability analysis by the finite element method
Heloisa Faggion1, Arthur Rossito2, Celso Mendonça3
1Hospital do Trabalhador, Curitiba, Brazil. heloisafaggion@gmail.com.
Purpose:
This study aimed to evaluate, using the finite element method (FEM), the biomechanical stability of a 3.5 mm cannulated screw for sacroiliac fixation of pelvic fractures in children aged ≤ 5 years.
Methods:
A computerized pediatric pelvic model was constructed from a CT scan of a 5-year-old patient. A Torode and Zieg Type IV fracture was simulated and fixed with a 3.5 mm cannulated screw positioned in the oblique S1 corridor. A bipedal support load condition was applied. For FEM analysis, two fixations were tested: one with a 3.5 mm screw (test) and another with a 7.0 mm screw (control). Visual and quantitative assessments of bone fragment displacement, total pelvic displacement, and maximum von Mises stress in the screws were performed.
Results:
Both constructs showed minimal fracture displacement and equivalent stability. The maximum von Mises stress in the 3.5 mm screw was 738 MPa, exceeding the steel alloy limits for plastic deformation and tensile strength, producing a small region of increased surface stress.
Conclusion:
In a FEM-based model, sacroiliac fixation with a 3.5 mm cannulated screw in the oblique S1 safety corridor provided stable fixation of pediatric pelvic fractures up to 5 years of age. Nevertheless, the stress observed could lead to screw failure. Therefore, strict postoperative load restriction is recommended.