Accidental reactivation of a retained pacemaker leading to competitive ventricular pacing between two pacemakers: A
Xiaomao Wu1, Zelin Li1, Lulu Xu1
1Zhejiang Hospital of Integrated Traditional Chinese and Western Medicine, Hangzhou 310003, Zhejiang, China.
Abstract:
Herein, we report a rare clinical case of competitive ventricular pacing caused by the accidental reactivation of a retained pacemaker two years after device replacement. Initially, two years following the implantation of the new pacemaker, the device demonstrated typical dual-chamber pacing. However, recent follow-up electrocardiography revealed two distinct QRS morphologies. Holter monitoring, imaging studies, and a review of prior procedural records confirmed that this phenomenon was caused by the unexpected reactivation of the retained pacemaker, leading to competitive ventricular pacing. A detailed analysis of the underlying causes of inadvertent reactivation led to the implementation of targeted interventions, which successfully resolved the issue of competitive ventricular pacing induced by dual pacemakers.
More Related Videos
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Mechanism of Cardiac Arrhythmias
Electrophysiology of Normal Cardiac Rhythm
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias