A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
A phantom study to achieve comparable 124I PET recovery and address scanner variability for dosimetry purposes in
Maaike Dotinga1, Lioe-Fee de Geus-Oei2, Petra Dibbets-Schneider3
1Department of Radiology, Section of Nuclear Medicine, Leiden University Medical Center, Leiden, the Netherlands; Department of Nuclear Medicine, Netherlands Cancer Institute, Amsterdam, the Netherlands.
Background:
Accurate quantification of iodine uptake is essential for performing pre-treatment dosimetry of 131I therapy after redifferentiation of radioiodine-refractory thyroid cancer. Standardized procedures for 124I PET/CT-based dosimetry are currently lacking. We aim to evaluate the relation between 18F and 124I imaging for two different PET/CT scanners, the effect of bias correction on recovery across scanners and investigate the impact of clinically-encountered background (BG)-to-lesion ratios on recovery correction.
Methods:
Cylindrical and NEMA body phantoms were scanned following vendor-recommended 124I acquisition using clinically representative activity concentrations (5.6-5.7 kBq/mL in cylindrical phantom, 45.1-59.2 kBq/mL in NEMA spheres) and BG-to-sphere ratios (∼1:125 to 1:infinity) using two digital PET/CT scanners (Philips Vereos and GE Healthcare Omni). Additionally, BG-to-sphere ratio ∼1:10 was acquired to compare 124I to the EARL 18F standards 1 (EARL1). Calibration accuracy and recovery coefficients (RCmax, RCmean) were compared between scanners with and without bias correction.
Results:
124I recovery was ∼20% higher for the Omni compared to the Vereos, showing calibration accuracies of 1.12-1.15 vs. 0.94, and RCmean reaching 0.86 vs. 0.69. After bias correction, RCmean was comparable between scanners (<1%) but below the lower limits of EARL1. A single fit for recovery correction (R2 = 0.97) was obtained for different BG-to-sphere ratios for both scanners as RCmean was comparable (p > 0.4).
Conclusion:
Vendor-recommended 124I acquisition and reconstruction leads to differences in quantification but can be compensated using a bias correction. Recovery correction is minimally affected by different BG-to-lesion ratios, suggesting that one RC curve is sufficient, simplifying 124I calibration procedures in studies requiring 124I quantification.