The Effect of Contrast Media Formulations with Different Iodine Preparation Concentrations at a Constant Iodine
Mathias Holtkamp1, Gregor Jost2, Luca Salhöfer1
1Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany.
Purpose:
The effect of iodine preparation concentration on vascular attenuation during computed tomography angiography (CTA) is a topic of controversy. Iodine-based contrast media (CM) formulations are used in different iodine preparation concentrations and, accordingly, their viscosity varies considerably. This study aimed to investigate the influence of CM formulations with different iodine preparation concentrations on image quality under conditions of a constant iodine dose and constant iodine delivery rate (IDR) in low-dose, low-kVp thoracoabdominal CTA using a standardized animal model.
Materials And Methods:
A cohort of six healthy Göttingen minipigs underwent three low-dose CTA scans (90 kV, CTDIvol = 1.7 ± 0.5 mGy) using three CM formulations with different iodine preparation concentrations (300, 370, and 400 mgI/mL). All formulations were injected at a constant IDR of 1.3 gI/s and at a constant total iodine dose of 252 mgI/kg in a randomized sequence. Quantitative image quality was assessed using contrast-to-noise ratio (CNR) and image similarity metrics. Subjective image quality was evaluated by two experienced radiologists using a 5-point Likert scale.
Results:
No significant differences in CNR were observed between the CM formulations (300 mgI/mL = 15.6 ± 2.8, 370 mgI/mL= 15.0 ± 3.8, 400 mgI/mL = 15.0 ± 4.9). The subjective image quality evaluations revealed no significant differences between the groups, except for a slight tendency towards higher ratings for the 300 mgI/mL formulation for sharpness (300 mgI/mL=5 (4.3-5), 370 mgI/mL = 4 (4-5), 400 mgI/mL = 4 (4-5)) and diagnostic acceptability (300 mgI/mL = 4 (3-4), 370 mgI/mL = 3.5 (3-4), 400 mgI/mL = 3.5 (2.3-4)).
Conclusion:
Under the investigated conditions of constant iodine delivery rate and constant iodine dose, CM formulations with iodine preparation concentrations ranging from 300 to 400 mgI/mL do not lead to significant differences in objective or subjective image quality in low-dose, low-kVp CTA. These findings represent a proof-of-concept under controlled experimental conditions.
Key Points:
· A standardized animal model allowed a controlled analysis of CM formulations with different iodine preparation concentrations.. · At a constant IDR and total iodine dose, image quality remained comparable across the investigated CM formulations.. · A higher iodine preparation concentration did not result in a measurable image quality benefit with a fixed IDR..
Citation Format:
· Holtkamp M, Jost G, Salhöfer L et al. The Effect of Contrast Media Formulations with Different Iodine Preparation Concentrations at a Constant Iodine Delivery Rate in Low-kV CT Angiography: An Experimental Animal Study. Rofo 2026; DOI 10.1055/a-2818-5244.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...