Influence of Tube Voltage, Tin Filtration, and Sex on Radiation Dose and Image Quality in Adult Chest Photon-Counting
Laura Valentina Klüner1, Sebastian Zensen1, Hannah Aus der Wiesche2
1Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany.
Purpose:
Photon-counting computed tomography (PCCT) offers new possibilities for optimizing image quality while reducing radiation dose. However, the influence of acquisition parameters such as tube voltage and spectral shaping with tin filtration on chest imaging remains insufficiently understood.
Materials And Methods:
Adult male and female anthropomorphic phantoms were scanned using a dual-source PCCT system across a broad range of image quality levels (IQL 1-100). Radiation dose and signal-to-noise ratio (SNR) were evaluated for protocols using 120 kVp and 140 kVp without tin filtration, and 100 kVp with tin filtration.
Results:
The male phantom required a significantly higher tube current and radiation dose than the female, reflecting sex-based attenuation differences. Increasing the tube voltage to 140 kVp raised the radiation dose with minimal SNR improvement. Tin filtration substantially reduced the dose despite a higher mAs, with the radiation dose and tube current plateauing near IQL 80. In the low-dose range (IQL 1-20), tin filtration enabled dose reduction without saturation. Slight SNR trade-offs between bone and mediastinum highlight the need for tissue-specific considerations when optimizing PCCT chest protocols.
Conclusion:
Sex- and tissue-specific differences with tin filtration and varying tube voltages underscore baseline differences in radiation dose and image quality behavior between representative adult body geometries. These baseline relationships provide a technical foundation for future patient-specific protocol optimization.
Key Points:
· A higher tube voltage increased the radiation dose with limited image quality benefit.. · Tin filtration enables substantial radiation dose reduction in low-dose adult chest PCCT.. · Tin filtration causes image quality and dose plateaus at image quality levels around 80.. · Baseline radiation dose and image quality relationships support future patient-specific chest PCCT optimization..
Citation Format:
· Klüner LV, Zensen S, aus der Wiesche H et al. Influence of Tube Voltage, Tin Filtration, and Sex on Radiation Dose and Image Quality in Adult Chest Photon-Counting CT: A Phantom-Based Signal-to-Noise-Based Analysis. Rofo 2026; DOI 10.1055/a-2829-0195.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography