The Quasi-Adiabatic Graphite Calorimeter for Absolute Dosimetry in Ultrahigh Dose Rate FLASH X-Ray Radiation Therapy
Ji Huang1, Guolong Zhang1, Kun Wang1
1Division of Ionizing Radiation Metrology, National Institute of Metrology (NIM), Beijing, China.
Purpose:
FLASH radiation therapy using high-energy x rays combines ultrahigh dose rate irradiation with the physical characteristics of high-energy x-ray beams, achieving a significant reduction in normal tissue biological damage while maintaining sufficient tissue penetration, thereby presenting great potential for clinical translation. However, the absence of a traceable absolute dosimetry method for FLASH x-ray beams remains a substantial limitation to its clinical implementation. This study aims to establish a quasi-adiabatic water-controlled probe-type graphite calorimeter for the absolute measurement of absorbed dose to water in the 10 MV FLASH x-ray beam, to address the current lack of a traceable dosimetry standard for x-ray FLASH radiation therapy.
Methods And Materials:
A probe-type graphite calorimeter was developed, employing thermally stabilized water as the thermal control medium to precisely regulate the thermal equilibrium of the graphite core. This quasi-adiabatic system is designed to facilitate accurate absolute dose measurements under ultrahigh dose rate conditions.
Results:
The results indicate that for a single irradiation with a total dose exceeding 2 Gy, the mean type A relative uncertainty, determined from 5 repeated measurements using the sample standard deviation, is less than 0.2%. By deriving the necessary correction factors for determining the absolute dose (ie, in Gy) of FLASH photon radiation therapy, the uncertainty in water absorbed dose measurement is determined to be 1.0% (k = 1).
Conclusions:
This study develops a probe-type graphite calorimeter for the absolute measurement of absorbed dose to water in 10 MV FLASH x-ray beams. The system is designed to address the current lack of a traceable dosimetric standard for x-ray FLASH radiation therapy, thereby supporting its clinical translation and application.