Validity and Reliability of Portable Blood Lactate Analyzers: A Systematic Review with COSMIN and GRADE Assessment
João Bruno1,2,3, Rohit Kumar Thapa4, Ângelo Santos5
1University of Coimbra, CIPER, Faculty of Sport Sciences and Physical Education, Coimbra, Portugal. joao_bruno97@hotmail.com.
Portable blood lactate analyzers (PBAs) show good reliability and validity for assessing performance. Standardized protocols and calibration are crucial for consistent results compared to lab methods.
Area of Science:
- Sports Medicine
- Biomedical Engineering
- Clinical Diagnostics
Background:
- Portable blood lactate analyzers (PBAs) are widely used in sports and clinical settings.
- Blood lactate is a key biomarker for performance and metabolic threshold assessment.
- Concerns exist regarding PBA accuracy, sampling, and calibration variability.
Purpose of the Study:
- Systematically review studies on the validity and reliability of commercial PBAs.
- Compare PBA performance against laboratory reference methods.
- Evaluate study quality and evidence certainty using COSMIN and GRADE frameworks.
Main Methods:
- Systematic literature search following PRISMA 2020 guidelines across major databases.
- Included 25 studies assessing PBA validity and reliability.
- Employed statistical measures (e.g., CCC, LoA, ICC, CV) and COSMIN/GRADE for quality assessment.
Main Results:
- PBAs demonstrated strong correlations (r=0.95-0.99) and high reliability (ICC>0.90).
- Systematic biases were noted at higher lactate concentrations.
- Thirteen studies exhibited high methodological quality.
Conclusions:
- PBAs offer strong reliability and moderate-to-high validity for lactate measurement.
- Standardized protocols and enhanced calibration are necessary for consistency with lab analyzers.