Evaluating the impact of the water content of hydrophobic intraocular lenses on visual quality
Jungmin Kim1,2, Young-Sik Yoo3,4, Jinheung Park5
1Department of Electrical and Computer Engineering, Inha University, 100, Inha-ro, Incheon, 22212 Republic of Korea.
Abstract:
The impact of the water content of hydrophobic intraocular lenses (IOLs) on visual quality remains unclear. This study investigated how the water content influenced photic phenomena. Optical bench studies were conducted to compare two hydrophobic trifocal IOLs: PanOptix® (low-water content IOL) and Clareon® PanOptix® (high-water content IOL). Modulation transfer function (MTF) measurements were used to evaluate optical quality, while point spread function (PSF) images were used to evaluate photic phenomena. To validate the optical bench results, a retrospective observational case series study was conducted using medical records for 85 eyes from 85 patients (PanOptix®, n = 38; Clareon® PanOptix®, n = 47). Postoperative visual outcomes were evaluated two months after the operation using a defocus curve for uncorrected visual acuity and contrast sensitivity (CS) with and without glare. Subjective satisfaction regarding spectacle independence and dysphotopsia was evaluated using a questionnaire. The optical bench study showed that the high-water content IOL exhibited a 6.16% smaller glare diameter in near-vision PSF analysis, and slightly higher MTF values at near distances. The clinical study supported these results, showing that the high-water content IOL provided better uncorrected visual acuity (P < 0.05) and spectacle independence (P = 0.016) at near distances. Additionally, the high-water content IOL showed higher contrast sensitivity with glare at 6.3° (P = 0.040) and 4° (P = 0.018). Hydrophobic IOLs with a higher water content improved contrast sensitivity and subjective satisfaction by reducing glistening and minimizing photic phenomena.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview
Ophthalmic Drug Delivery Systems
Angle Closure Glaucoma: Treatment
Focusing of Light in the Eye
Quality of Water