Automation of Polysaccharide Quantification: A Rapid High-Throughput Assay Enabled by Liquid Handling Technology
Samuel Nicacio1, Winston Umakanth Balasundaram1, Aboli Bhingarkar1
1Quality Control, Inventprise, Inc., Redmond, WA 98052, USA.
Biotech (Basel (Switzerland))
|March 27, 2026
Summary
A new automated anthrone assay using a liquid handling system (LHS) offers a corrosion-mitigated solution for polysaccharide quantitation. This method enables unattended, high-throughput analysis of pneumococcal polysaccharides, improving efficiency and accuracy.
Area of Science:
- Analytical Chemistry
- Biochemistry
- Vaccine Development
Background:
- Traditional anthrone assays are common for polysaccharide quantitation but are labor-intensive and unsuitable for high-throughput needs.
- The corrosive nature of the anthrone-sulfuric acid reagent poses challenges for automation.
- Accurate polysaccharide quantification is crucial for vaccine development and quality control.
Purpose of the Study:
- To develop and validate a fully automated, corrosion-mitigated anthrone assay using a liquid handling system (LHS).
- To assess the accuracy, precision, and linearity of the automated assay compared to the manual method.
- To validate the automated assay according to ICH Q2 (R2) guidelines for reliable polysaccharide quantitation.
Main Methods:
- Development of an automated anthrone assay integrated with a liquid handling system (LHS).
- Quantification of 23 pneumococcal (PNU) polysaccharide serotypes using both manual and automated anthrone assays.
- Validation of the automated LHS method according to ICH Q2 (R2) guidelines, assessing accuracy, precision, and linearity.
Main Results:
- The automated anthrone LHS method achieved accuracy between 81-115% and precision with a coefficient of variation <7.0%.
- Linearity was demonstrated in the 30-650 µg/mL range with R² ≥ 0.993.
- The automated assay processes eight samples per hour, supports overnight batch operation, and automates all steps from pipetting to data export.
Conclusions:
- The developed automated anthrone assay is the first fully unattended and corrosion-mitigated method validated under ICH Q2 (R2) for polysaccharide quantitation.
- This automated LHS method provides an efficient, accurate, and precise alternative to manual anthrone assays for high-throughput analysis.
- The system facilitates overnight batch processing, significantly enhancing workflow efficiency in vaccine component analysis.