Purpose:
The present study aimed to assess the radiation dose associated with low-dose (LD) CT pelvimetry compared with conventional radiography and to evaluate the adequacy of the resulting image quality.
Methods:
The absorbed dose was measured using thermoluminescent dosimeters positioned in an anthropomorphic female phantom, including uterine locations, to estimate the fetal dose. Conventional radiographic pelvimetry and LD-CT pelvimetry were performed using clinically implemented protocols. Effective dose was calculated using Monte Carlo-based modeling applying acquisition parameters and retrospective patient dose registry data. Image quality of LD-CT pelvimetry was independently evaluated in 14 consecutive clinical cases using a four-point ordinal scale.
Results:
LD-CT pelvimetry reduced the mean absorbed pelvic dose by approximately 50% compared with conventional pelvimetry (0.18 vs. 0.39 mGy) and decreased estimated fetal dose by 40% (0.21 vs. 0.37 mGy). These estimates were based on standardized single acquisitions and did not incorporate additional radiation from retakes commonly observed in conventional practice. CT demonstrated substantially more homogeneous dose distribution, whereas conventional pelvimetry exhibited marked heterogeneity with peak values up to 2.3 mGy. The maternal effective dose was lower for LD-CT (0.16 mSv) than for conventional pelvimetry (0.36 mSv); inclusion of retakes increased the conventional effective dose to 0.71 mSv. All CT examinations were diagnostically adequate, and no recalls were required.
Conclusions:
Optimized low-dose CT pelvimetry significantly reduces radiation dose compared with conventional radiographic pelvimetry while maintaining reliable diagnostic image quality. These results support the clinical adoption of CT-based pelvimetry as a dose-efficient and reproducible alternative to conventional techniques.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT