Validation of 12-lead Smartphone-based spandan ultra 12L electrocardiogram durations and amplitudes analysis as
Chandra Mohan1, Kunal Gururani1, Anurag Rawat1
1Department of Cardiology, Himalayan Institute of Medical Sciences, Swami Rama Himalayan University, Dehradun, Uttarakhand, India.
Abstract:
Background. Electrocardiography (ECG) is a valuable tool for diagnosing cardiac disorders, with a 12-lead ECG being the gold standard. However, the traditional 12-lead ECG machine is hospital-based, which restricts access. The Smartphone-based 12-Lead ECG (Spandan Ultra 12L ECG) is a portable, simultaneous 12-lead ECG device designed for real-time cardiac monitoring. This study compares its accuracy with a Standard 12-lead ECG system (Philips PageWriter ECG).Objective. To evaluate the accuracy and reliability of the Smartphone-based 12-Lead ECG device in measuring ECG interval, amplitude, and duration measurements as compared to the Standard 12-lead ECG.Methods. A prospective, single-blinded, cross-sectional study was conducted in a hospital ECG room in Dehradun, India. 405 participants aged ≥ 20 years in ECG recordings using the both the ECG machine. After applying the exclusion criteria, 75 participants were excluded, and 330 participants' ECG recordings were analyzed. Statistical analyses included paired t-tests to compare significant differences, Pearson correlation coefficients (PCC) to assess linear association, and Bland-Altman (BA) analysis to evaluate the agreement between the two devices.Results. Twelve-lead Standard (n = 330) and Smartphone-based 12-Lead (n = 330) ECG waveforms from 330 participants (mean age 52.14 years); 169 (51.21%) males were analyzed. No statistically significant differences were found between both the ECGs for all measured parameters, with p-values ranging from 0.090 for P-wave duration to 0.965 for Q-wave amplitude. PCC showed high linear correlation between the both the ECGs across ECG intervals (0.818-0.938), amplitudes (>0.92), and durations (0.700-0.944). BA analysis confirmed minimal bias, with limits of agreement (LOA) within clinically acceptable ranges.Conclusion. The Smartphone-based 12-Lead ECG device showed high accuracy and reliability in the evaluation of ECG parameters, as compared to the Standard 12-lead ECG system. Its portability, smartphone compatibility, and ease of use make it a promising diagnostic tool for regular clinical applications and timely ECG assessment.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias V: Evaluating Dysrhythmias