A Short-Lived Isotope Counting System (SLICS) to measure reaction cross sections producing light, beta-decaying
A K Schwemmlein1, M Yuly2, S J Padalino3
1Laboratory for Laser Energetics, 250 East River Rd, Rochester, New York 14623, USA.
Abstract:
A new activation detector optimized for short-lived isotopes was designed, installed, and tested on OMEGA. The Short-Lived Isotope Counting System (SLICS) can be positioned as close as 35 cm from the OMEGA target chamber center in a ten-inch-manipulator. Stable target samples can be mounted on the front end of the detector assembly to undergo triton-, deuteron-, or neutron-induced reactions that produce short-lived reaction products that beta decay. The emitted betas are detected in a phoswich configuration to eliminate the gamma background. The detector is optimized for isotopes with half-lives from 150 ms to 20 s, filling a gap in OMEGA's current activation detectors for isotopes with half-lives >5 min. In a pilot study, SLICS achieved an efficiency of 24.5% ± 2.0% for counting 8Li, with a total of 505.8 ± 4.4 qualifying events. Using the existing triton laser-ion acceleration platform on OMEGA, this detector can be used to measure rarely studied tritium-induced reactions on light nuclei with an estimated uncertainty of 12%. The design of the detector and its performance during a first experiment employing deuterium as a surrogate will be discussed.
Related Concept Videos
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Nuclear Stability
To hold positively charged protons together in the...
Radioactive Decay and Radiometric Dating
Isotopes and Radioisotopes
An isotope containing more...
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.