Related Experiment Video
Updated: Apr 4, 2026

Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
Continuous-flow ventilation with VENTIJET in moderate ARDS: a pilot safety and feasibility study
Lucía Picazo Moreno1,2, Andrea Castellví-Font1,2, Maria Acer Puig1
1Critical Care Department, Hospital del Mar de Barcelona. Critical Illness Research Group (GREPAC), Hospital del Mar Research Institute (HMRI), Passeig Marítim, 25-29, 08003, Barcelona, Spain.
Background:
Acute respiratory distress syndrome (ARDS) remains associated with high mortality and ventilator-induced lung injury. Continuous-flow ventilation (CFV) may reduce cyclic strain, but faced technical limitations. VENTIJET is a novel nozzle-based CFV ventilator designed for safe use in critically ill adults. This pilot study evaluated its feasibility and safety in patients with moderate ARDS.
Methods:
We conducted a prospective, single-center pilot safety and feasibility study in intubated adults with moderate ARDS and stable oxygenation. After a 1-h baseline period with volume-controlled ventilation (Puritan Bennett™ 840), patients received up to 24 h of VENTIJET ventilation, then returned to conventional ventilation. Primary endpoints were feasibility and safety; secondary exploratory endpoints included gas exchange and respiratory mechanics.
Results:
Fourteen patients were enrolled. VENTIJET ventilation was successfully delivered in all participants without protocol interruptions or device-related serious adverse events. After 1 h, PaO₂/FiO₂ was 200 mm Hg compared to 192 mm Hg at baseline (p = 0.090). Over 24 h, oxygenation and respiratory system compliance showed modest improvement, with stable carbon dioxide (CO₂) elimination, transpulmonary pressures, and hemodynamics.
Conclusions:
In patients with moderate ARDS, VENTIJET CFV was feasible and safe, maintaining stable gas exchange with modest improvements in respiratory mechanics. These findings support further evaluation in larger, controlled trials, including the full spectrum of ARDS. Trial registration ClinicalTrials.gov Identifier NCT07121257.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: