Impact of partial volume correction on radiomics reproducibility in theranostic SPECT/CT imaging
Mohammad Saber Azimi1,2, Maryam Cheraghi3, Mohammad Ali Ghodsi Rad4
1Doctoral School of Applied Informatics and Applied Mathematics, Óbuda University, Budapest, Hungary.
Background:
Radiomics has shown potential for quantitative characterization of tumors in molecular imaging; however, its clinical translation in theranostic 1 7 7Lu SPECT/CT remains limited due to poor robustness of extracted features to reconstruction variability and partial volume effects. Establishing reproducible radiomics biomarkers across correction strategies is therefore a prerequisite for reliable clinical modeling and treatment monitoring.
Purpose:
This study aimed to evaluate radiomics feature reproducibility, defined as the stability of feature values across different partial volume correction (PVC) strategies and reconstruction settings, in clinical 1 7 7Lu SPECT/CT imaging. In addition, we explored two volumetric shape-based indices, the metastasis-to-liver ratio (MLR) and metastasis-to-spare liver ratio (MSLR), as surrogate markers of hepatic metastatic burden in the theranostic treatment setting.
Methods:
In 13 patients (40 scans) treated with 177Lu, 837 radiomics features were extracted from 11 abdominal regions and metastases on SPECT/CT using original and wavelet-decomposed images across four bin widths (50-200). Two post-reconstruction PVC methods, namely Richardson-Lucy (RL) and Reblurred Van Cittert (RVC), were applied. Feature reproducibility was quantified using two complementary metrics: the intraclass correlation coefficient (ICC) to assess feature-level stability across PVC strategies, and the concordance correlation coefficient (CCC) to evaluate pairwise agreement and systematic bias among reconstruction methods. Visual image quality assessments were independently performed by two experienced nuclear medicine specialists in a blinded setting. Exploratory metastatic tumor burden was assessed descriptively using 3D shape-based MLR and MSLR indices.
Results:
Low-frequency wavelet decomposition (LLL-wavelet) and original features showed the highest reproducibility (ICC ≥ 0.90 in >95% of liver and metastasis features at BW50), whereas high-frequency features and larger bin widths demonstrated reduced stability. CCC analysis revealed excellent agreement between RL and RVC (≥0.95 in major organs at BW50-100), while agreement with uncorrected SPECT (no PVC) was consistently lower, especially for high-frequency features. RL achieved higher visual scores in sharpness and contrast (p < 0.01), with good inter-reader agreement supporting the consistency of these assessments. MLR/MSLR demonstrated inter-patient variability and were explored descriptively as indices of metastatic liver burden.
Conclusions:
Reproducibility in theranostic SPECT radiomics is highly feature- and organ-dependent and is further influenced by scanner-specific factors and reconstruction protocols, which remain critical for real-world clinical translation. RL and RVC showed stronger mutual agreements than each with uncorrected SPECT. Importantly, only RVC translated visual improvements into enhanced feature-level reproducibility, while RL provided the most consistent overall balance of reproducibility and image quality, supporting its role as the preferred PVC strategy for clinical and modeling applications. Robust radiomics feature selection as well as standardized reproducible PVC strategies are essential to generate methodological harmonization for future clinical translation and to support integration of radiomics analyses into personalized SPECT theranostics.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...