Safety of 3 Tesla Magnetic Resonance Imaging in Active Peripheral Nerve Field Stimulation Device for Facial Pain

Clement T Chow1,2,3,4, Can Sarica1,2, Benson Yang5

  • 1Division of Neurosurgery, Department of Surgery, University of Toronto, Toronto, Ontario, Canada.

Abstract

Insights

MRI scans are safe for patients with active peripheral nerve field stimulation (PNFS) devices for facial pain. This study confirmed safety and feasibility, enabling functional MRI (fMRI) use in this population.

Area of Science:

  • Neuromodulation
  • Medical Imaging
  • Pain Management

Background:

  • Peripheral nerve field stimulation (PNFS) uses electrical stimulation for facial pain.
  • Functional MRI (fMRI) assesses central effects of neuromodulation but its safety with PNFS is unknown.
  • This limits clinical MRI use and fMRI application in PNFS patients.

Purpose of the Study:

  • Evaluate MRI safety and feasibility for active SCS implants in PNFS patients.
  • Define parameters for concurrent blood-oxygenation-level-dependent (BOLD) fMRI acquisition.
  • Establish safe imaging protocols for PNFS patients.

Main Methods:

  • In vitro safety testing using a 3D-printed phantom with an SCS implant.
  • Evaluated head specific absorption rate (SAR) and B1+rms for temperature changes.
  • Validated findings in a single patient using structural and BOLD fMRI sequences.

Main Results:

  • fMRI during active PNFS is safe with temperature increases below 2°C.
  • Head SAR showed a stronger relationship with heating than B1+rms.
  • A patient underwent MRI without adverse events, achieving successful fMRI acquisition.

Conclusions:

  • Phantom testing confirmed MRI safety and feasibility with active SCS for facial PNFS.
  • Specific conditions were identified for safe MRI acquisition.
  • Context-specific safety evaluations are crucial for medical device MRI compatibility.