Comparison of Fingertip Hemoglobin Value with Venous Hemoglobin Value in Bleeding Patients Followed at the Emergency
Background:
Hemoglobin's pivotal role in human health is monitored through hospital-based complete blood count (CBC) by utilizing automated methods like electrical impedance and photon scattering. This study explored bedside fingertip hemogram and hematocrit analysis' reliability versus CBC´s in emergency department patients, crucial for hemorrhagic condition management.
Methods:
The study included emergency department patients requiring hemogram monitoring. Participant data, CBC, and fingertip hemogram and hematocrit result values from venous blood using a conventional device were compared during monitoring.
Results:
In this study, 88 participants (63.6% female, mean age 47.68 ± 19.80) were assessed for hemoglobin and hematocrit levels using both fingertip and conventional methods. Fingertip hemogram measurements showed a significant correlation with conventional CBC devices, yet tended to be 1.20 g/dL lower. Similarly, fingertip hematocrit measurements correlated positively but were 3.59% lower compared to conventional measurements.
Conclusions:
Bedside fingertip devices for hemogram measurement in emergency departments provide time-efficient and resource-saving advantages.
Related Concept Videos
Venous Thrombosis IV: Nursing Management
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...