Related Experiment Video
Updated: Apr 11, 2026

In Silico Clinical Trials for Cardiovascular Disease
Published on: May 27, 2022
Cardiac Multifrequency Time-harmonic Elastography: In Silico Validation and In Vivo Application
Stefan Klemmer Chandía1, Brunhilde Wellge2,3, Gina Barzen3,4
1Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Department of Radiology, Germany, Berlin.
Optimized cardiac time-harmonic elastography (THE) accurately detects abnormal diastolic stiffness in transthyretin amyloidosis (ATTRwt) patients. This portable ultrasound method offers a rapid, low-cost approach for diagnosing cardiomyopathies.
Area of Science:
- Cardiovascular Imaging
- Medical Ultrasound
- Biomedical Engineering
Background:
- Myocardial stiffness, particularly diastolic stiffness, is a key indicator of cardiac dysfunction.
- Transthyretin amyloidosis (ATTRwt) is a progressive disease characterized by abnormal myocardial stiffening.
- Accurate, non-invasive methods are needed to assess myocardial stiffness and diagnose conditions like ATTRwt.
Purpose of the Study:
- To introduce and validate an optimized multifrequency cardiac time-harmonic elastography (THE) technique.
- To assess THE's ability to detect abnormal diastolic stiffness using portable drivers and multi-heartbeat acquisitions.
- To investigate THE's diagnostic performance in patients with cardiac wild-type transthyretin amyloidosis (ATTRwt) compared to healthy controls.
Main Methods:
- Cardiac THE was performed using a portable driver inducing continuous vibrations at 60, 70, and 80 Hz.
- Shear wave speed (SWS) maps, a proxy for myocardial stiffness, were acquired over 4 seconds.
- In silico simulations using CT-data-based anatomies validated the technique and optimized parameters. Diagnostic performance was assessed using ROC analysis.
Main Results:
- The study included 10 healthy controls and 11 patients with ATTRwt.
- Cardiac THE accurately detected ATTRwt with an AUC of 1.00 (95% CI: 0.94-1.00).
- In silico SWS estimations showed strong agreement with ground truth (r²=0.94) across various stiffnesses and thicknesses.
Conclusions:
- Cardiac THE accuracy was quantified for the first time in silico and in vivo, including an ATTRwt cohort.
- Multifrequency shear waves from portable drivers, encoded in standard ultrasound, offer rapid, low-cost contrast.
- This technique aids in detecting cardiomyopathies such as amyloidosis by assessing myocardial stiffness.
More Related Videos
10:53Ultrasonic Assessment of Myocardial Microstructure
Published on: January 14, 2014
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010