Threshold-Dependent Discordance Between STRATIFY JCV™ and IMMUNOWELL™ Assays in Relapsing Multiple Sclerosis
Clara Grazia Chisari1,2, Pina Crimì3,4, Giulio Spampinato3,4
1Department of Medical, Surgical Sciences and Advanced Technologies "GF Ingrassia", University of Catania, Via S. Sofia 78, 95123, Catania, Italy. clara.chisari@unict.it.
Anti-JC virus (JCV) antibody testing for progressive multifocal leukoencephalopathy (PML) risk stratification shows good agreement between STRATIFY JCV™ and IMMUNOWELL™ assays. Concordance improves significantly when using clinical risk categories, supporting interchangeability for natalizumab-treated patients.
Area of Science:
- Immunology
- Neurology
- Clinical Diagnostics
Background:
- Anti-JC virus (JCV) antibody testing is crucial for stratifying progressive multifocal leukoencephalopathy (PML) risk in natalizumab-treated patients.
- STRATIFY JCV™ is the established reference assay, while IMMUNOWELL™ is gaining traction with biosimilar use.
- Understanding inter-assay variability is key for accurate risk assessment.
Purpose of the Study:
- To compare the concordance between STRATIFY JCV™ and IMMUNOWELL™ assays for anti-JCV antibody testing.
- To determine if discordance stems from biological differences or interpretation thresholds.
- To evaluate assay performance within clinically relevant PML risk categories.
Main Methods:
- A single-center observational study involving 250 relapsing multiple sclerosis (RMS) patients.
- Paired, same-day testing using STRATIFY JCV™ and IMMUNOWELL™ assays.
- Analysis of agreement using manufacturer cutoffs and clinical PML risk thresholds, with correlation and Bland-Altman analyses.
Main Results:
- Concordance was 78.0% (κ=0.55) using analytical cutoffs, with discordance in borderline ranges.
- Concordance increased to 93.2% (κ=0.73) when applying clinical risk thresholds.
- Strong correlation (r=0.922) and minimal mean bias (+0.06) were observed, with deviations mainly at lower index values.
Conclusions:
- Inter-assay discordance between STRATIFY JCV™ and IMMUNOWELL™ is primarily driven by interpretation thresholds.
- Applying clinical PML risk categories significantly improves assay concordance.
- The findings support the interchangeability of these assays for routine monitoring in the biosimilar era, with caution advised for borderline results.
More Related Videos
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Immunodeficiency Diseases
There are three main causes of immunodeficiency...
Myasthenia Gravis: Diagnostic Tests
The edrophonium test is a diagnostic tool for myasthenia gravis. It involves...
Immunoprecipitation
Chromatin Immunoprecipitation
Chromatin immunoprecipitation, also known as ChIP, is used to study protein-DNA or...
Cytomegalovirus Disease
Autoimmune Disorders
Concept and Mechanism of Autoimmune Diseases
The immune...