Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Analytical and Clinical Performance of the Atellica HEMA 580 Haematology Analyzer: A Comparison With the ADVIA 2120i
Hyeongyu Lee1, Jooyoung Cho1, Gilsung Yoo1
1Department of Laboratory Medicine, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea.
Introduction:
The Atellica HEMA 580 is a novel haematology analyser providing eight-part leukocyte differential and optical platelet (PLT) counting. We compared its analytical and clinical performance with that of the ADVIA 2120i.
Methods:
A total of 796 residual whole blood specimens were analysed. Carryover, linearity and precision were assessed. Inter-instrument comparison was performed for complete blood counts and five-part differential (5-Diff) parameters using Passing-Bablok regression and Bland-Altman analyses. Additional regression against manual microscopy was conducted for flagged samples, including PLT, reticulocytes and common 5-Diff parameters.
Results:
Carryover was < 1% for all parameters. Linearity was excellent (R2 ≥ 0.99). Precision was within acceptable limits; coefficient of variant was < 5% for most parameters, except optical PLT, lymphocytes, monocytes, eosinophils and nucleated red blood cells under specific conditions. Inter-instrument comparisons revealed remarkably high correlations (r ≥ 0.9) for most parameters, whereas mean corpuscular haemoglobin concentration (MCHC) (r = 0.429) and basophils (r = 0.627) showed weaker concordance. Passing-Bablok regression indicated minimal systematic or proportional bias for most parameters. Bland-Altman analysis showed that > 95% of results were within the 95% limits of agreement, except white blood cells, mean corpuscular volume and PLT (94.7%-94.8%). Comparison with microscopy confirmed good concordance for PLT and reticulocytes; however, lymphocytes and monocytes had proportional and systematic biases.
Conclusion:
The Atellica HEMA 580 demonstrated reliable analytical performance and strong agreement with the ADVIA 2120i. This supports its suitability as a routine haematology analyser, providing accurate and clinically comparable results.