Continuous bed motion versus step-and-shoot acquisition in LAFOV PET/CT: insights from multi-phantom and patient
P M Linder1, W Lan1, E Calderón1
1Department of Nuclear Medicine and Clinical Molecular Imaging, University hospital Tuebingen, Tuebingen, Germany.
Background:
Continuous bed motion (CBM) allows flexible extension of the scan range compared to conventional step‑and‑shoot (S&S) acquisition but has not yet been evaluated in long axial field‑of‑view (LAFOV) PET/CT. This study systematically assessed the impact of CBM on image quality, noise, and quantitative performance in the Biograph Vision Quadra LAFOV PET/CT using multi‑phantom and patient scans compared to S&S.
Methods:
A uniform tube phantom and a NEMA IEC phantom, positioned centrally and off-centre, were scanned across bed speeds (2.8-50 mm/s), sensitivity modes and scan ranges (106 and 150 cm) to evaluate image uniformity, axial count profiles, noise and contrast recovery coefficients (CRC). Ten oncological patients receiving [18F]PSMA-1007 or [18F]FDG underwent sequential CBM (2.8 mm/s, 378 s) and S&S (300 s) scans. Image noise, net true counts, and liver and lesion SUV values were compared using paired statistics and Bland-Altman analysis, along with PSMA expression scores.
Results:
For comparable count statistics and image noise, CBM required a prolonged acquisition (378 s) to match the S&S (300 s) protocol, resulting in comparable image quality for phantoms and patients. CRC and image uniformity were preserved across all evaluated conditions, even at the FOV's axial edge (50.5 cm) for 8.4 mm/s (22 mm sphere: CRC 76% S&S vs. 71% CBM). In patient scans, minor differences in axial count profiles, net true counts, and SUV values (SUVmean bias - 0.1 (liver) and - 0.8 (lesions)) did not affect clinical scores.
Conclusions:
The prolonged CBM protocol provides image quality and quantitative performance comparable to S&S in LAFOV PET/CT. While the reconstructed image range remains constrained by CT coverage, the patient scan comparison with 106 cm scan range, together with extended range phantom measurements, indicate that CBM can support scan range extension beyond 106 cm without compromising diagnostic accuracy.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...