Related Experiment Video
Updated: Apr 15, 2026

Conversion of a Capture ELISA to a Luminex xMAP Assay using a Multiplex Antibody Screening Method
Published on: July 6, 2012
Evaluating Random-Access CLIA for Anti-TNF Drug Monitoring: a Comparison with Traditional ELISA
The novel i-TRACK10 system offers accurate therapeutic drug monitoring for anti-TNF biologics, showing high agreement with ELISA methods. This chemiluminescence immunoassay provides a reliable and efficient alternative for clinical labs.
Area of Science:
- Clinical Chemistry
- Immunology
- Pharmacology
Background:
- Accurate therapeutic drug monitoring (TDM) of anti-tumor necrosis factor (anti-TNF) biologics like infliximab and adalimumab is crucial for managing inflammatory diseases.
- Monitoring corresponding antibodies is essential for optimizing patient treatment efficacy and safety.
Purpose of the Study:
- To evaluate the analytical and clinical performance of the new i-TRACK10 chemiluminescence immunoassay (CLIA) system.
- To compare the i-TRACK10 system with established LISA Tracker enzyme-linked immunosorbent assay (ELISA) kits for TDM of anti-TNF agents.
Main Methods:
- Analyzed 200 clinical serum samples for infliximab, adalimumab, and their antibodies using both i-TRACK10 and ELISA platforms.
- Assessed method agreement using Deming regression, Bland-Altman analysis, and Lin's concordance correlation, adhering to CLSI EP09-A3 guidelines.
- Calculated positive/negative agreement and precision using sample sets and controls.
Main Results:
- The i-TRACK10 showed strong association (Pearson's r: 0.91-0.95) and high positive/negative agreement (>95%) with ELISA.
- Concordance varied (Lin's CCC: 0.75-0.93) due to method biases, confirmed by Deming regression and Bland-Altman analysis.
- Demonstrated low intra- and inter-assay variability (CV < 5%), rapid turnaround, expanded ranges, and random-access operation.
Conclusions:
- The i-TRACK10 analyzer is a reliable and accurate alternative to ELISA for anti-TNF biologic TDM.
- Its enhanced automation and performance characteristics facilitate integration into routine clinical laboratory workflows.
- The system offers an efficient solution for optimizing anti-TNF therapy management.
More Related Videos
07:52Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
16:07Determination of the Relative Potency of an Anti-TNF Monoclonal Antibody mAb by Neutralizing TNF Using an In Vitro Bioanalytical Method
Published on: September 16, 2017