Impact of (defined) hydration on renal shear wave elastography
Antonia Sowa1, Miriam Kesselmeier2, Ulrike John-Kroegel3
1Diagnostic and Interventional Radiology, Section of Pediatric Radiology, Jena University Hospital, Jena, Germany.
Summary
Renal shear wave elastography (SWE) in healthy adults shows hydration impacts stiffness differently based on baseline elasticity. Standardized protocols are crucial for consistent, interpretable kidney elastography results.
Area of Science:
- Medical Imaging
- Ultrasound Technology
- Renal Physiology
Background:
- Elastography is a noninvasive ultrasound method for tissue mechanical property assessment.
- Renal elastography is not yet standard clinical practice.
- Understanding factors influencing renal elastography is essential for its clinical adoption.
Purpose of the Study:
- To analyze renal elastography values in healthy adults.
- To investigate the influence of hydration status on renal elastography.
- To identify the most suitable anatomical location for renal elastography measurements.
Main Methods:
- Shear wave elastography (SWE) performed on 20 healthy adults before and after 500 mL water intake.
- Measurements taken from upper pole, lower pole, and hilar regions of the kidney.
- Hilar region identified as the optimal site based on measurement variability and accessibility.
Main Results:
- Fluid intake's effect on tissue stiffness varied with baseline stiffness; higher baseline stiffness showed smaller increases or greater decreases.
- No significant influence of participant characteristics (age, sex, etc.) on SWE parameters was observed.
- The hilar region was determined to be the most suitable site for measurement.
Conclusions:
- Hydration's impact on renal SWE parameters is modulated by the baseline elasticity of the renal parenchyma.
- Standardized protocols are necessary for comparable and interpretable renal elastography outcomes.
- Baseline elastographic values influence the changes induced by fluid intake, highlighting the need for standardized assessment.
Related Concept Videos
Imaging Studies II: Ultrasonography
803
IntroductionUltrasonography, or renal ultrasound, is a noninvasive medical imaging technique that uses high-frequency sound waves to visualize the kidneys, ureters, bladder, and surrounding tissues.Indications for Urinary System UltrasonographyUrinary system ultrasonography is indicated in various clinical scenarios, such as:Kidney Stones (Urolithiasis): To detect and monitor the size and presence of kidney or urinary tract stones.Hydronephrosis: To assess the dilation of the renal pelvis and...
803
Ultrasound II: Endoscopic Ultrasound and FibroScan
1.2K
Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
Endoscopic Ultrasound (EUS):
1.2K