Comparison of photon-counting CT with energy-integrating CT in temporal bone imaging: an anthropomorphic phantom
David Pellby1,2, Malin Vestin-Fredriksson3, Anton Rönnblom3
1Department of Medical Imaging and Function, Skåne University Hospital, Lund/Malmö, Sweden.
None:
BackgroundPhoton-counting computed tomography (PCCT) is an emerging technique under evaluation for temporal bone imaging. Its capacity for radiation dose reduction while maintaining diagnostic image quality (IQ) requires further study.PurposeTo evaluate how PCCT compares to energy-integrating CT (EICT) in temporal bone imaging in terms of IQ and visualization of anatomical landmarks, at a clinically relevant, reduced radiation dose.Material and MethodsAn anthropomorphic phantom with exchangeable fresh frozen human temporal bone specimens was scanned. Nine specimens were scanned on a PCCT system with clinical protocol settings, followed by stepwise dose reductions for one specimen. Ten specimens were scanned on a EICT system with clinical protocol settings. Three readers assessed overall IQ and visualization of 10 important anatomical landmarks on a 4-point scale.ResultsPCCT was preferred by readers for most landmarks and ranked higher in overall IQ compared with EICT. At clinical protocol, PCCT provided statistically significant better results for all landmarks and overall IQ compared to EICT with dose reduction of almost 70%. With stepwise dose reduction PCCT maintained good to excellent IQ for most landmarks, suboptimal for the ossicular chain landmarks at lower doses, and overall poor image quality at the lowest dose level. PCCT at clinical and several reduced dose levels demonstrated statistically significant difference in overall image quality and visualization of landmarks, compared to EICT (P <0.05).ConclusionPCCT provides improved diagnostic image quality compared to EICT at clinical protocols, with lowered radiation dose, indicating its suitability for temporal bone imaging.
More Related Videos
13:16Computed Tomography and Optical Imaging of Osteogenesis-angiogenesis Coupling to Assess Integration of Cranial Bone Autografts and Allografts
Published on: December 22, 2015
07:28Thinned-skull Cortical Window Technique for In Vivo Optical Coherence Tomography Imaging
Published on: November 19, 2012
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...