Related Experiment Video
Updated: Apr 19, 2026

Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Real-time torque behavior of reciprocating nickel-titanium instruments using different irrigating solutions
Uğur Dursun1, Mevlüt Sinan Ocak2
1Batman University, Faculty of Dentistry, Department of Endodontics, Batman, Turkey.
Instrument design, not irrigating solutions, primarily impacts mechanical stress during root canal preparation. Different file systems and irrigant combinations showed varied torque responses, with specific instruments generating higher torque.
Area of Science:
- Endodontics
- Dental Materials Science
- Biomaterials Engineering
Background:
- Reciprocating nickel-titanium instruments are used in root canal preparation.
- Mechanical stress during preparation can compromise instrument safety.
- The influence of irrigating solutions on operative torque is not well understood.
Purpose of the Study:
- To investigate the impact of different irrigating solutions on torque during root canal shaping.
- To compare torque generation across various reciprocating single-file systems.
- To analyze the interaction between irrigant type and file system on torque.
Main Methods:
- Extracted human mandibular premolars with oval canals were used.
- Three reciprocating file systems were tested with sodium hypochlorite, ethylenediaminetetraacetic acid, or saline.
- Torque-controlled motors recorded real-time operative torque and preparation time.
Main Results:
- Irrigating solutions alone did not significantly affect torque or preparation time.
- A significant interaction between file system and irrigant type was observed.
- The T-Endo MUST system generated significantly higher mean and peak torque compared to other instruments.
Conclusions:
- Instrument design characteristics appear to be more critical than irrigating solutions in influencing mechanical stress.
- The specific combination of file system and irrigant affects torque response.
- Understanding these interactions is crucial for optimizing root canal preparation safety and efficiency.