Collision energy consistency between analyte and internal standard: A critical factor for accurate quantification of
Xiaoli Ma1, Wei Luo1, Yixuan Cai1
1Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing, 100730, China.
Abstract:
Simultaneous quantification of steroid hormones with vastly differing physiological concentrations using LC-MS/MS often necessitating compromised collision energy (CE) settings, potentially introducing inaccuracy. This study investigates the critical impact of CE consistency between analytes and their stable isotope-labeled internal standards (SIL-IS) on quantification accuracy using a triple quadrupole LC-MS/MS system. We analyzed six hormones in 61 plasma samples using a novel dual-CE acquisition method, monitoring SIL-IS response at both low and optimal CE. The method was reapplied to 70 patient samples following collision cell replacement to assess instrument dependency. Significant instrument-specific bias (15-92%) were observed for high-concentration analytes when CE settings between native analytes and SIL-IS were mismatched (LCE-HCE). These biases were markedly amplified on a degraded instrument. Synchronizing CE settings (LCE-LCE) or utilizing consistently optimized isotopic transitions restored accuracy, reducing mean bias to below 15% post-optimization. The instrument in optimal condition showed minimal CE-dependent bias. Inconsistent CE application is a hidden source of quantitative inaccuracy in multi-analyte steroid panels, exacerbated by suboptimal instrument conditions. While these findings provide important methodological insights, further studies on different LC-MS/MS platforms are warranted to establish their generalizability. We strongly recommend verification of CE consistency during method validation and instrument troubleshooting.
More Related Videos
08:53Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species
Published on: May 7, 2018
11:25Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
High-Resolution Mass Spectrometry (HRMS)
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...