Multispectral Real-time Fluorescence Imaging for Intraoperative Detection of the Sentinel Lymph Node in Gynecologic Oncology
Published on: October 20, 2010
Fluorescence changes at flow cytometric analysis of samples from sentinel lymph nodes excised with indocyanine green
Alessandra Ubiali1, Elisa Maria Gariboldi1, Damiano Stefanello1
1Dipartimento di Medicina Veterinaria e Scienze Animali, Università degli Studi di Milano, Lodi, Italy.
Introduction:
Flow cytometry (FC) has been used recently to assess percentages of infiltration by mast cells in sentinel lymph nodes (SLN) of dogs with mast cell tumors (MCT). SLN mapping often includes the use of different dyes such as methylene blue (MB) and indocyanine green (ICG), which might influence fluorescence assessment by FC. This study aimed to assess whether the color given by mapping dyes might affect the baseline fluorescence of SLN aspirates for FC.
Methods:
Baseline fluorescence was calculated as Median Fluorescence Intensity (MFI) in 4 channels (FL1, FL2, FL3, FL4, respectively corresponding to 530/30 nm, 585/40 nm, >670 nm, 660/20 nm) with a cytometer equipped with two lasers (488 nm and 638 nm) with constant setting and compensation. SLN aspirates were suspended in RPMI medium. They were classified based on results of mapping techniques in 4 dye classes (blue/fluorescent, fluorescent/non-blue, blue/non-fluorescent, non-blue/non-fluorescent), and possible differences in MFI values among SLN dye classes were assessed.
Results:
Thirty-five SLNs from 17 dogs were assessed. Considering dye classes, 13 were blue/fluorescent, 16 were fluorescent/non-blue, 4 were blue/non-fluorescent, 2 were non-blue/non-fluorescent. In all fluorescence channels, except FL1, MFI varied among SLN dye classes. Blue/non-fluorescent SLNs showed the highest fluorescence, followed by blue/fluorescent, fluorescent/non-blue, and non-blue/non-fluorescent.
Discussion:
These results suggest that the use of dyes for SLN mapping may introduce a relevant bias when MFI is quantitatively assessed via FC.