Related Experiment Video
Updated: Apr 21, 2026

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Percutaneous salvage of a failed 14F suture-mediated femoral closure using dual collagen-mediated vascular closure
Estefania Rivera-Pedraza1, Luis A Santiago-Sulsona2,3, José L Ortiz-Fullana2,3
1University of Puerto Rico School of Medicine, San Juan, Puerto Rico.
Abstract:
Transcatheter aortic valve replacement is performed via percutaneous femoral access; however, vascular complications can occur owing to the use of large-bore sheaths. We describe a novel technique for managing a failed suture-mediated percutaneous closure of a 14F arteriotomy using two collagen-based closure devices. We achieved hemostasis without loss of access and avoided an open surgical repair. This highlights the utility of dual MYNX (MYNX vascular closure device) as a rescue strategy for closing large diameter arteriotomies. Our experience suggests that this technique may be a salvage option in cases of suture-mediated closure failure, avoiding open repair.
More Related Videos
20:33A Case Series of Successful Abdominal Closure Utilizing a Novel Technique Combining a Mechanical Closure System with a Biologic Xenograft that Accelerates Wound Healing
Published on: July 4, 2019
06:36A Novel Tenorrhaphy Suture Technique with Tissue Engineered Collagen Graft to Repair Large Tendon Defects
Published on: December 10, 2021