Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Middle Ear Pressure Assessment to Detect the Stapedius Reflex: A Prospective Observational Feasibility Study
José Luis Vargas Luna1, Johannes Proksch1, Katharina Geißler2
1MED-EL Medical Electronics, Innsbruck, Austria.
Objective:
To evaluate the feasibility of detecting stapedius reflex (SR) through direct middle ear pressure monitoring during contralateral acoustic stimulation.
Study Design:
Prospective, observational, single-centre feasibility study.
Setting:
Tertiary referral centre; intraoperative setting during tympanotomy procedures.
Patients:
Eleven adult patients (≥18 y) with unilateral sudden sensorineural hearing loss undergoing transtympanic dexamethasone instillation, with an elicitable SR through contralateral stimulation, were included in this study. Patients with SR thresholds >100 dB HL were excluded.
Intervention S:
A pressure transducer was inserted intraoperatively into the middle ear. A tympanometer device on the contralateral side apply acoustic stimulation and monitored the SR on that side. Stimulation was ECG-synchronized, and filtering was applied to the pressure signal to enhance detection.
Main Outcome Measures:
Detection of SR-related pressure changes using the pressure transducer, compared with tympanometry-based detection.
Results:
Out of 32 stimulus combinations, 24 elicited confirmed stapedius reflexes, while 8 did not. Among the 24 reflex-positive cases, the pressure-based method correctly identified reflex-related pressure changes in 20 instances, yielding a detection accuracy of 83.3% for confirmed reflexes. In the remaining 4 reflex-positive cases, the system failed to detect a response. In all 8 cases where no reflex was elicited, the system also did not detect any pressure changes, indicating no false positives.
Conclusions:
Middle ear pressure monitoring is a feasible method for detecting SR. This approach demonstrates feasibility under intraoperative conditions and may, with further development and validation, support future objective reflex assessment in cochlear implant users.
Related Concept Videos
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Physical Assessment of the Respiratory Tract III: Percussion
Percussion in Respiratory Assessment
Percussion evaluates underlying tissue composition with audible and tactile vibrations,...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.
Assessment of blood pressure in brachial artery(two-step method)
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities: