Reduction of CT number location dependency using photon-counting detector CT and virtual monoenergetic imaging
Maryam Sadeghian1,2, Joseph Swicklik1, Cynthia H McCollough1,2
1Department of Radiology, Mayo Clinic, Rochester, MN, United States of America.
None:
Objective.Computed tomography (CT) number accuracy is critical in radiotherapy planning. This study aimed to investigate the potential of photon-counting-detector (PCD) CT to reduce location dependency of CT numbers compared to energy-integrated-detector (EID) CT across imaging modes, tissue types, and patient sizes.Approach.Four inserts of typical tissue types: cortical bone, liver, lung, and adipose were placed inside a body size (40 × 30 cm2) multi-energy CT phantom (Gammex). To simulate a larger patient, tissue-equivalent material (Superflab) was added to increase phantom lateral width to 50 cm. Both phantoms were scanned on a PCD-CT (NAEOTOM Alpha, Siemens), a dual-source EID-CT (Force, Siemens; EID-A), and a dual-layer EID-CT (7500 Spectral, Philips; EID-B). All scans were performed with 120 kV, except dual-energy mode of EID-A (100/Sn150 kV, Sn: Tin filter). Volume CT dose index (CTDIvol) was matched across scanners. Each insert was scanned in seven locations: iso-center and six peripheral locations. In addition to single-energy images, 70 keV virtual-monoenergetic images (VMIs) were reconstructed. CT numbers were measured using circular regions of interest. Mean CT number, standard deviation (std), coefficient of variation (CV), max-to-min and central-to-peripheral differences were quantified. CT number variation across seven locations was assessed using the Friedman test and Bonferroni corrected pairwise Wilcoxon signed-rank tests.Main results.PCD-VMI demonstrated lower CT number locational variation compared to EID-CTs among most tissue inserts and phantom sizes. For cortical bone in the standard phantom, PCD-VMI achieved std of 3.31 HU, showing 70.3% reduction compared to single-energy PCD (11.15 HU) and lower than single-energy EID-A (9.75 HU), single-energy EID-B (10.93 HU), EID-A VMI (12.65 HU), and EID-B VMI (4.10 HU). PCD-VMI CV remained below 1% and 3% for small and large phantoms, respectively. Increasing phantom size increased CT number variations, with lowest differences for PCD-VMI. PCD-VMI showed significantly lower CT number locational variation than all other scans across most inserts and phantom sizes.Significance.PCD-VMI reduces CT number locational dependency compared to most imaging modes. The greatest improvements were observed for high-Zmaterials and larger patients. PCD-VMIs are promising for radiation therapy applications.
More Related Videos
12:24Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Determination of Crystal Structures