Determination of a safe radiation distance from a 16-slice multi-detector computed tomography (MDCT) scan machine
A I Hocson1, N D P Concepcion1
1Institute of Radiology, St. Luke's Medical Center, Global City, Taguig, Metro Manila Philippines; e-mails: aihocson@gmail.com; npconcepcion@stlukes.com.ph.
Abstract:
The objective of this study is to determine the location with the lowest radiation levels inside the CT scan room while performing examinations using a 16-slice multi-detector computed tomography (MDCT) scanner. This was an analytical, experimental study taking radiation dose measurements using a dosimeter placed one location at a time using a clock-face pattern with 1-m intervals during a simulated CT-guided biopsy procedure. A total of 51 dose measurements were taken. The lowest radiation dose at 1 m is at the 9:00 position, then at the 3:00 position. The lowest radiation dose at 2 m is at the 9:00 position. The doses in these three positions do not have a statistical difference, but are significantly lower than the rest of the doses in other positions. The dose of radiation in a CT scan room is dependent on the distance and direction of the radiation source. If medical personnel need to stay inside the room during procedures, the safer distance from the radiation source was 1 and 2 m away at the 9:00 position and 1 m away at the 3:00 position, which are on the sides of the CT scanner.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography