Validation of a freely available Eclipse script for automatic calculation of EQD2 dose distribution in external beam
Juan-Francisco Calvo-Ortega1, Coral Laosa-Bello2, Sandra Moragues-Femenía2
1Servicio de Oncología Radioterápica, Hospital Quirónsalud Barcelona, Barcelona, Spain; Servicio de Oncología Radioterápica, Hospital Quirónsalud Málaga, Málaga, Spain.
None:
To validate a script designed to streamline equivalent dose in 2 Gy fractions (EQD2) calculations for external beam radiotherapy. The VarianESAPI-EQD2Converter script for EQD2 calculation was evaluated for the Eclipse v. 16.1 treatment planning system. Seven anatomical sites were included (head and neck, lung, liver, breast, prostate, abdomen, and brain), and 3 cases per site were randomly selected from our clinical database. For each plan, the values of the D2%, D50% and D98% metrics of the planning target volume (PTV), and D2%, D50%, and D0.1cc of the organs-at-risk (OARs) were retrieved and converted into EQD2 values using the EQD2 formula ("true values"). Then, the script was executed for each plan following these 2 approaches: (1) the alpha-beta values of the PTV and OARs were simultaneously entered (approach 1); (2) only the external patient contour ("body") structure was considered, and the script was run several times by entering the alpha-beta values of the PTV and OARs each time. Validation of the script was performed by comparing the EQD2 values computed by the script for the Dx%/D0.1cc metrics with the respective true values. A 2-tailed t-test statistical analysis was done (alpha = 0.05). Nonstatistically significant differences were found between the 2 approaches, but unacceptably high differences (up to 57.5%) were found using approach 1. By contrast, low absolute differences (mostly < 0.5%) were found with approach 2. The VarianESAPI-EQD2Converter script with approach 2 is a reliable method to automate the EQD2 calculation process for Eclipse users.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Radioactivity and Nuclear Equations
A nuclide of an element has a specific number of protons and...
Biological Effects of Radiation
Calculations of Electric Potential II
Consider a...
Calculation of Electric Flux
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses