Current Practices and Clinical Implementation of Electronic Portal Imaging Device-based Dosimetry in Indian
Raghavendra Hajare1,2, Rohit Avinash Vadgaonkar1, Shanmukhappa B Kaginelli2
1Department of Radiation Oncology, Homi Bhabha Cancer Hospital and Research Centre, Visakhapatnam, Andhra Pradesh, India.
Purpose:
This study aims to assess the current prevalence, utilization patterns, and operational challenges of electronic portal imaging device (EPID)-based dosimetry systems in Indian radiotherapy centres.
Materials And Methods:
A multiple-choice survey was distributed through Google Forms to medical physicists in India. The four-section survey covered treatment techniques, EPID availability and dosimetry, alternative detectors, and future challenges. Responses were limited to one per institution, with follow-up reminders.
Results:
A total of 237 responses were received with a response rate of 62.2%. Among these, 73% of institutions currently use EPID-based dosimetry systems. Of these users, 42.2% rely on EPID for daily quality assurance (QA), and 62.4% of institutions use EPID for more than 50% of their total QA workflow. The primary tasks performed using EPID systems include patient-specific QA (PSQA) for intensity modulated radiation therapy (100%), periodic machine QA (58.4%), PSQA for stereotactic radiation therapy/stereotactic body radiation therapy (48.6%), machine commissioning (18.5%), and in vivo dosimetry (10.4%). Among institutions not currently using EPID due to nonavailability, 89.1% expressed willingness to adopt it for PSQA in the future. Major challenges in implementing EPID dosimetry were software limitations (34.2%) and maintenance and calibration issues (35.0%). After adopting EPID dosimetry, 79.8% of institutions observed improvements compared to conventional detectors. In addition, 93.7% of institutions believed that EPID dosimetry implementation would reduce physicist workload through automated software and resource optimization.
Conclusion:
This study reveals the widespread use of EPID systems and highlights that implementation offers certain advantages. However, issues such as software limitations, high initial cost, and the need for affordable or free resources remain barriers to wider adoption.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiation: Applications
The average...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Absorption of Radiation
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET