Computed Tomography Dose Index Measurements in Wide-beam Computed Tomography System
Sawwanee Asavaphatiboon1, Suwadee Payomthip2, Prapa Sodkokkruad1
1Department of Radiology, Faculty of Medicine, Ramathibodi Hospital, Mahidol University, Bangkok, Thailand.
Abstract:
Computed tomography (CT) produces cross-sectional images for medical diagnosis; however, in wide cone-beam CT, conventional CT dose index (CTDI) measurements using a 100-mm ionization chamber (IC) underestimate dose for larger beam widths. This study evaluated and compared CTDI measurements in wide beam using different methods for wide-beam CT, following International Atomic Energy Agency Human Health Report No. 5. Measurements were performed on an Aquilion™ ONE CT scanner (160 mm beam width) using 100-and 300-mm ICs (Radcal 10 × 6-3CT and PTW TM30017). CTDIfree-in-air and weighted CTDI (CTDIw) were obtained under brain and abdomen protocols, both in free air and phantom conditions, at the Advanced Diagnostic Imaging Center, Faculty of Medicine Ramathibodi Hospital, Thailand. For free-in-air measurements at 80 mm beam width, CTDI100 air (no-step) was slightly smaller than CTDI300 air with percentage differences of -1.27% and -1.94%, while CTDI100air (two-step) showed +6.79% and +6.48% differences for brain and abdomen protocols, respectively. At 160 mm beam width, CTDI100 air (no-step) was significantly lower due to incomplete dose coverage, whereas two-and three-step methods yielded slightly higher values. For CTDIw at 80 mm, percentage differences were -17.06% and -15.05% (no-step), -3.29% and +1.88% (two-step), and -8.94% and -3.45% (calculated two-step) for brain and abdomen, respectively. At 160 mm, CTDI100w (no-step) was 41.27% and 36.38% lower than CTDI300W, while CTDI100w (three-step) exceeded CTDI300w due to scattering or overlap. Overall, CTDI measurements using a 100 mm IC underestimate dose for beam widths >NT +40 mm. The two-step technique is sufficient for wide-beam CT dosimetry measurements.
More Related Videos
12:24Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...