Multi-institutional Comparison of Novel O-ring Linac Cone Beam Computed Tomography with Fan Beam Computed Tomography
Luis Agulles-Pedrós1,2, Robert Lee MacDonald1, Amanda J Cherpak1
1Department of Medical Physics, Nova Scotia Health, Halifax, Nova Scotia, Canada.
Aim:
This work presents a multi-institutional image comparison between novel O-ring linac cone-beam computed tomography (CBCT) and fan-beam computed tomography simulator (FBCT), among five institutions.
Materials And Methods:
A phantom was sent to five institutions with Ethos and Halcyon units equipped with HyperSight CBCT (HS-CBCT). HyperSight used a 125 kVp/176 mAs protocol and an iterative-CBCT reconstruction algorithm with scatter correction. FBCT imaging protocols used 125 kVp and exposure from 176 to 379 mAs.
Results:
Linear fitting of relative electron density versus Hounsfield unit (RED-vs-HU) curves was determined for RED ≤ 1 and RED ≥ 1. The contrast was evaluated with regard to solid water. Noise and contrast-to-noise ratio (CNR) were evaluated with and without exposure normalization. The RED-vs-HU curve for RED ≤ 1 of FBCT shows a 4% greater slope than HS-CBCT (P = 0.024), while for the RED ≥ 1 case, the FBCT slope is 15% larger compared to (P = 0.00004). HS-CBCT has a larger contrast as RED differs from 1, although at RED > 1, the difference is significant, up to 14%. However, noise is larger for HS-CBCT, especially for RED > 1, up to 10 times. Even with exposure correction, noise is more significant for HS-CBCT, which leads to a CNR ten times smaller than FBCT (RED = 1.78).
Conclusion:
This multi-institutional analysis of HS-CBCT showed reduced slopes of the RED-vs-HU curves compared to FBCT, which results in a greater sensitivity of the HS-CBCT to RED changes. HS-CBCT shows better contrast performance, although the broader beam leads to noise up to 10 times larger noise, even with the scatter correction. The CNR of the FBCT is up to one order of magnitude greater than that for HS-CBCT.
More Related Videos
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...