Finite element analysis of internal fixation with clavicular hook plate vs. adjustable loop plate for Neer Type IIB
Liwen Zheng1, Baopeng Li1, Xi Wang1
1Chengdu BOE Hospital, Chengdu, China.
Objective:
To compare the biomechanical stability of clavicular hook plates vs. adjustable strap plates in treating Neer Type IIB distal clavicle fractures, providing evidence for clinical procedure selection.
Methods:
Based on CT data from a healthy volunteer, a three-dimensional finite element model of the shoulder joint-including bones, ligaments, and internal implants-was constructed using Mimics, Geomagic, and SOLIDWORKS software. Model A simulated fixation with a clavicular hook plate, while Model B simulated fixation with an adjustable strap plate. Muscle loads (simulating holding a 0.5 kg teacup) and boundary conditions were set based on literature. Static mechanical analysis was performed in ANSYS Workbench. Observed metrics included Total Deformation at the path node on the clavicular upper surface (assessing stability) and Equivalent Stress (assessing refracture risk), with statistical comparisons conducted.
Results:
Under simulated loading, both models provided effective fixation. Internal fixation material Stresses remained below their respective yield strengths (titanium alloy: 830 MPa; suture: 1,208.87 ± 164.87 MPa), indicating no fracture risk. Model A (hook plate) exhibited a smaller maximum Deformation on the clavicle (0.54 mm) compared to Model B (0.79 mm), but the mean Deformation at path nodes showed no statistically significant difference (P > 0.05). The mean Equivalent Stress at the clavicular path node for Model A (3.42 MPa) was greater than that for Model B (2.87 MPa), with a statistically significant difference (P < 0.05). The maximum Stress in Model A was concentrated at the hook bend (221.15 MPa), potentially generating localized high pressure on the subacromial surface; the maximum Stress in Model B was located at the midpoint of the strap connection line (278.67 MPa).
Conclusion:
Finite element analysis indicates that both the clavicular hook plate and the adjustable strap plate provide adequate biomechanical stability for Neer Type IIB distal clavicle fractures. While both demonstrate comparable fixation strength, they exhibit distinct Stress distribution characteristics. Considering the risk of complications such as subacromial impingement associated with clavicle hook plates, while arthroscopic strap plate techniques can simultaneously address concomitant shoulder injuries without requiring secondary surgery for removal, the latter may offer greater overall advantages.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026
Related Concept Videos
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Statically Indeterminate Problem Solving
Stability of structures
Internal Loadings in Structural Members: Problem Solving
To illustrate this, let's consider a beam OC of 5 kN, inclined at an angle of 53.13° with the horizontal and supported at both ends. Determine the internal...