X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Shielding of the Trunk During Extremity Imaging Reduces Radiation Exposure in Pediatric Patients
Simon Blanchard1, Steven Zhang2, William Huffman2
1Orthopedic Center, The Children's Hospital of Philadelphia.
Background:
Although pediatric patients are particularly sensitive to the effects of ionizing radiation, there are no established protocols for shielding the trunk during extremity orthopaedic procedures. However, some suggest that shielding not be used on the patient as it may increase radiation exposure. In this study, we aimed to quantify the effect of lead shielding on the trunk at 3 critical anatomic locations during upper-extremity imaging.
Methods:
A mannequin (Laerdal Medical) approximating the size of a 5-year-old was positioned supine with the upper-extremity extended over a hand table. Exposures of the forearm were performed with a fluoroscopy machine (GE Healthcare) at 110 kVp and 4.5 mA for 10 seconds. A dosimeter (Mirion Technologies) was used to assess radiation exposure at the neck/thyroid, chest/breast, and groin. Readings were assessed with the image intensifier in both the above and the below configuration, with and without a radiation shield placed in different configurations over the trunk. Each configuration was replicated 5 times.
Results:
Without shielding, the neck/thyroid received 110.5 and 132.7 mSv, the breast/chest received 125.8 and 118.8 mSv, and the groin received 44.9 and 46.7 mSv with the receiver below and above, respectively. These values were significantly higher than those with an anterior shield (42.9 and 80.8, 37.0 and 68.8, and 12.9 and 32.3 mSv). The addition of a side lead significantly reduced radiation exposure compared with anterior shielding alone in most locations (19.4 and 95.8, 24.0 and 46.7, 0.0 and 0.0).
Conclusion:
Radiation shields placed anteriorly over a child significantly reduced radiation exposure in 3 anatomic areas. The addition of a side wrap-around shield further reduced exposure.
Clinical Relevance:
Although fluoroscopy is commonly used in extremity procedures, there is no standard guideline for radiation shielding in pediatric orthopaedic procedures. This data suggests that physicians should strongly consider placing shielding to create a radiation shadow that blocks scattered radiation from the imaged extremity. While the mannequin was the size of a child, these recommendations may be translatable to adult patients as well.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...