Heated and Humidified Oxygen in Critical Care Transport: Evaluating the Use of Aerogen Pro-X Nebulizer as an
Maria Karas1, Trace Choulat2, Jamie Persten3
1Department of Biomedical Engineering, College of Engineering, University of Florida, Gainesville, FL.
Objective:
Heated, humidified oxygen therapy is commonly used to support hypoxemic, nonintubated patients. Conventional high-flow nasal cannula (HFNC) systems are effective in hospital settings but are often large and less practical for use during critical care transport. Portable aerosol-generating devices offer a compact, transport-friendly alternative; however, a performance gap exists in their ability to maintain appropriate temperature and humidity under varying conditions compared with HFNC. We compared the effectiveness of a heated humidification device, Hamilton H900 (H900), with that of an aerosol-generating device, Aerogen Pro-X (Aerogen).
Methods:
The Aerogen was tested with a Hamilton T1 ventilator set to "high-flow oxygen device mode," as used in emergency responses. Using a hygrometer, we measured humidity and temperature outputs at 3 different set flow rates and 6 different set fractions of inspired oxygen (FiO2) levels. Each condition was tested the following 3 times: baseline establishment, Aerogen, and H900.
Results:
Across all conditions, the H900 delivered significantly higher temperatures than the Aerogen, with an estimated marginal mean of 38.56°C ± 0.34°C (101.41°F ± 2.61°F) compared with 19.62°C ± 0.34°C (67.32°F ± 2.61°F) for the Aerogen (95% confidence interval [CI]). In contrast, the Aerogen delivered significantly higher relative humidity (RH) than the H900 at 85.9% ± 0.9% compared with 57.7% ± 0.9% for the H900 (95% CI); however, its absolute humidity was lower than that of the H900 (14.53 mg/L vs. 27.42 mg/L for the H900). For both devices, RH decreased as FiO2 and flow rate increased (P < .001). At the highest settings (FiO2 100%, 60 liters per minute), RH declined to 71.3% ± 3.8% for the Aerogen and 42.3% ± 3.8% for the H900 (95% CI).
Conclusion:
The H900 maintained near-physiological temperatures, whereas the Aerogen demonstrated near-physiological RH. Although each device performed well within a specific domain, neither independently provided guideline-compliant humidification and temperature control across tested conditions. These findings support the development of a portable HFNC system that integrates efficient humidification with active temperature control for use in transport settings.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...