Related Experiment Video
Updated: Jul 4, 2026

Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Combined Snare and Buddy-Balloon Strategy for Failed Valve Crossing in Valve-in-Valve Transcatheter Aortic Valve
Hira Sarfraz1, Muhammad U Javed2, Parth K Parikh3
1Internal Medicine Residency Program, Eastern Virginia Medical School, Macon & Joan Brock Virginia Health Sciences at Old Dominion University, Norfolk, Virginia, USA.
Background:
Retrograde valve crossing during valve-in-valve transcatheter aortic valve replacement (ViV TAVR) may fail despite modern delivery systems.
Case Summary:
A 67-year-old man with a degenerated 27-mm Magna Ease bioprosthesis and unfavorable aortic geometry (kinked ascending aorta at a prior surgical graft site) underwent ViV TAVR, where retrograde valve crossing failed despite multiple stiff guidewires and snare traction alone. Successful delivery of a 29-mm Evolut FX+ was ultimately achieved by combining gooseneck-snare traction for coaxial realignment with a partially inflated Tyshak buddy balloon acting as a mechanical ramp across the surgical valve frame.
Discussion:
This case demonstrates a structured, stepwise approach to managing failed valve crossing in ViV TAVR, highlighting the mechanical roles of snare traction and balloon-mediated coaxial alignment.
Take-Home Message:
A combined snare and buddy-balloon strategy can facilitate valve delivery when conventional retrograde crossing fails during ViV TAVR.