Impact of automatic exposure control on radiation dose and detectability in dual-source and fast kV switching
Kosuke Matsubara1, Yoshinori Ogawa2, Ryo Yoshikawa2
1Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Ishikawa, Japan. matsuk@mhs.mp.kanazawa-u.ac.jp.
Abstract:
We evaluated radiation dose and detectability changes with automatic exposure control (AEC) according to object size in dual-source (DS) and fast kV switching (FS) dual-energy computed tomography (DECT). A phantom with five section diameters (16-36 cm) was scanned using different AEC settings (DS: Quality Reference mAs [QRmAs] 300-700; FS: Noise Index [NI] 8-12). Volume CT dose index (CTDIvol) and detectability index (d') for iodine were measured. Clinical CTDIvol data from 40 to 80 kg patients undergoing liver dynamic DECT were retrospectively analyzed. In DS-DECT, CTDIvol increased slightly with section diameter but plateaued at QRmAs 600-700 for 31-36 cm (31 cm: 24.1 mGy; 36 cm: 22.5-22.7 mGy), and d' decreased for larger sections. Clinical CTDIvol did not differ significantly among weight groups (40-<50 kg: 21.5 mGy; 50-<60 kg: 22.2 mGy; 60-<70 kg: 22.8 mGy; mean; p = 0.13). In FS-DECT, CTDIvol and d' varied with NI and section diameter: for the 26-cm section, CTDIvol ranged from 15.0 to 30.8 mGy and d' from 37.5 to 59.0; for 36-cm section, CTDIvol was 39.9 mGy and d' 24.0-27.8, with smaller variations than single-energy CT (SECT). Clinical CTDIvol increased with patient weight up to 70 kg (40-<50 kg: 20.3 mGy, 50-<60 kg: 25.8 mGy, 60-<70 kg: 29.2 mGy; mean; p < 0.05). AEC behavior in DECT differs from SECT, causing variations in dose and detectability. Appropriate AEC settings in DECT can achieve image quality comparable to SECT.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System V: CT