Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Photon counting detector CT contrast agent-reduced transcatheter aortic valve reconstruction planning: a comparative
Yannik C Layer1, Alexander Isaak2, Narine Mesropyan2
1Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Venusberg-Campus 1, 53127, Bonn, Germany. yannik.layer@ukbonn.de.
Objectives:
Continuous efforts are made to reduce contrast media, improving patient safety, reducing environmental risks, and addressing recurring supply shortages. The aim of this study was to evaluate contrast agent-reduced CT protocols for transcatheter aortic valve reconstruction (TAVR) planning in photon counting detector CT (PCDCT).
Materials And Methods:
162 BMI-matched examinations with standard dose contrast media (SCD; 80 mL; Iohexol 300 mg/mL; 81 examinations) and reduced contrast media dose (RCD; 50 mL; 81 examinations) for TAVR planning on a PCDCT were included in this retrospective monocentric study. Virtual monoenergetic reconstructions (VMI) at 70 keV, 60 keV and 50 keV of contrast agent-reduced examinations were compared with polyenergetic images. Quantitatively, regions-of-interest (ROIs) were placed in the abdominal aorta, iliac bifurcation, femoral artery, left ventricle and trapezius muscles. Signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR) were calculated. Qualitatively, diagnostic quality and contrast were assessed on a visual grading scale of 1 (non-diagnostic) - 5 (excellent) and contrast agent dose was estimated.
Results:
Averaged, SNR and CNR decreased by 8.71% and 16.78%, respectively, on PCDCT with reduced contrast dose (RCD vs. SCD; both p < 0.001). VMI50keV increased SNR by 44.10% (p < 0.001) and CNR by 52.73% (p < 0.001) compared with SCD. In the ascending aorta, SNR increased from 19.80 ± 6.24 (SCD) to 35.78 ± 13.20 (RCDVMI50keV) and CNR from 18.84 ± 7.78 to 29.77 ± 16.70. Median contrast intensity was 5 for SCD, 4 for RCDCR, and 5 for RCDVMI50keV.
Conclusion:
The diagnostic efficacy of TAVR planning assessment using PCDCT with minimized contrast agent dosing is preserved, presenting a practical approach to conserve contrast media.
Key Points:
Question The aim of the study was to implement a PCDCT-adapted contrast media dose protocol to reduce contrast agent volume at sufficient diagnostic quality. Findings PCDCT enables substantial contrast dose reduction for TAVR planning with maintained diagnostic image quality. Low-keV virtual monoenergetic image reconstructions compensate for the reduced iodine concentration. Clinical relevance The study demonstrates the potential of contrast media reduction of PCD-CT in clinical routine. This can benefit patients with renal impairment, for example, and reduce the negative effects of iodinated contrast media on the environment.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT