Closing the evidence gap: Capillary high-sensitivity troponin I verification in an African population using point of

Maseti Amvuyele Ann1, Jacob Doreen1, Khoza Siyabonga1

  • 1Department of Chemical Pathology, Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, South Africa; National Health Laboratory Service, Johannesburg, South Africa.

Insights

Point-of-care testing using capillary blood for high-sensitivity cardiac troponin (hs-cTnI) shows good agreement with lab tests. This supports its use for rapid acute myocardial infarction (AMI) diagnosis in resource-limited areas.

Area of Science:

  • Clinical Diagnostics
  • Cardiovascular Medicine
  • Point-of-Care Testing

Background:

  • High-sensitivity cardiac troponin (hs-cTnI) assays are crucial for diagnosing acute myocardial infarction (AMI).
  • Point-of-care (POC) platforms using capillary blood offer potential for faster results and improved access in resource-limited settings.
  • This study assessed the performance of capillary hs-cTnI POC sampling in an African population.

Purpose of the Study:

  • To evaluate the analytical performance and clinical concordance of capillary hs-cTnI POC testing against central laboratory assays.
  • To determine the feasibility of using capillary blood for rapid AMI assessment in resource-limited emergency departments.

Main Methods:

  • A cross-sectional method comparison study was conducted using paired capillary and venous samples from patients with suspected AMI.
  • Measurements were performed on a POC platform using capillary blood and a central laboratory assay using serum and plasma.
  • Agreement was assessed using Bland Altman analysis, and diagnostic concordance was determined at sex-specific 99th percentile thresholds.

Main Results:

  • Capillary hs-cTnI demonstrated moderate correlation with both plasma (rs=0.788) and serum (rs=0.791) assays.
  • Overall agreement at the 99th percentile threshold was high (92.1%), with substantial inter-method agreement (κ >0.75).
  • The capillary method showed acceptable bias compared to the serum assay, meeting predefined criteria.

Conclusions:

  • Capillary whole-blood hs-cTnI testing shows acceptable analytical agreement and high clinical concordance with central laboratory measurements.
  • This supports the implementation of capillary hs-cTnI POC testing to improve rapid AMI assessment in resource-limited emergency settings.
Abstract