Intraoperative continuous noninvasive hemoglobin monitoring in patients undergoing thoracic surgery
Mona Mohamed Mogahed1, Taha Saad Alnoamani2, Mohamed Shafik Elkahwagy3
1Associate Professor of Anesthesiology, Surgical ICU and Pain Medicine department, Faculty of Medicine Tanta University, Tanta, Egypt. monamogahedfr@hotmail.com.
Background:
This study aims to assess the agreement between noninvasive hemoglobin (SpHb) and invasive hemoglobin (InvHb) monitoring in guiding blood transfusions during thoracic procedures.
Methods:
This study was not interventional but assessed the agreement between SpHb and InvHb monitoring in guiding blood transfusions during thoracic procedures. Data were collected from 80 patients ≥ 18 years older who were candidates for thoracic surgeries. Continuous SpHb monitoring was conducted. Simultaneous (InvHb) samples were collected for validation at predefined perioperative intervals: preoperative baseline, immediately after induction, 1, 2, 3, and 4 h after incision, at the end of surgery, and in the recovery room. Agreement was assessed using Bland-Altman analysis, and correlation was evaluated using Pearson coefficients. The primary outcome measure was the correlation between SpHb and InvHb.
Results:
There was a significant positive correlation between SpHb and InvHb at preoperative period, immediately after induction of anesthesia, 1 h, 2 h, 3 h, 4 h after start of surgery, at the end of surgery and after recovery of the patient (Pearson's r = 0.949772, 0.940262, 0.915292, 0.981438, 0.989623, 0.984838, 0.965235 and 0.933264). Bland-Altman analysis between SpHb and InvHb demonstrated low bias with good limits of agreement at the same interval points [Mean bias (limits of agreement): 0.56 (-0.3 to 1.4), 0.54 (-0.38 to 1.5), 0.67 (-0.37 to 1.7), 0.52 (-0.14 to 1.8), 0.48 (0.12 to 1.1), 0.6 (-0.001 to 1.2) g/dl].
Conclusion:
In patients undergoing thoracic surgery, there is a strong correlation and acceptable agreement between SpHb monitoring using Masimo pulse co-oximetry and invasive hemoglobin measurements.
Related Concept Videos
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...