Scanning resolution influences texture features in intraoral radiographs
Bianca Costa Goncalves1, Pedro Santos Beckmann2, Victoria Geisa Brito de Oliveira2
1Institute of Science and Technology, Department of Diagnosis and Surgery, São Paulo State University (UNESP), Engenheiro Francisco José Longo Avenue, 777, Jardim São Dimas, São José dos Campos, São Paulo, 12245-000, Brazil. bianca.goncalves@unesp.br.
Lower scanning resolutions of photostimulable phosphor (PSP) plates create digital radiographic images with less uniformity. Texture analysis (TA) showed reduced homogeneity at lower resolutions, impacting image quality regardless of object thickness.
Area of Science:
- Medical Physics
- Radiology
- Image Analysis
Background:
- Digital radiography utilizes photostimulable phosphor (PSP) plates for image acquisition.
- Spatial scanning resolution is a critical parameter affecting digital image quality.
- Texture analysis (TA) provides quantitative measures of image texture and homogeneity.
Purpose of the Study:
- To investigate the impact of varying spatial scanning resolutions of PSP plates on digital radiographic images.
- To assess image texture characteristics using TA across different scanning resolutions.
Main Methods:
- Digital radiographs of an aluminum step wedge were acquired at varying exposure times.
- Images were scanned at multiple spatial resolutions (10, 20, 25, 40 lp/mm).
- Texture features were extracted using gray-level co-occurrence matrix (GLCM) and analyzed statistically.
Main Results:
- Texture analysis revealed significant differences in image texture parameters across spatial resolutions.
- Lower scanning resolutions (10 and 20 lp/mm) generally resulted in higher texture parameter values.
- These effects were observed independent of object thickness and, to some extent, exposure time.
Conclusions:
- Lower spatial scanning resolutions on PSP plates lead to reduced image homogeneity and uniformity.
- Texture analysis is a valuable tool for quantifying the effects of scanning resolution on digital radiographic images.
- Optimizing scanning resolution is crucial for maintaining diagnostic image quality in PSP systems.
Related Concept Videos
X-ray Imaging
Three-Dimensional Microscopy in Microbiology
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...